| Technical Parameter | DS-161 |
|---|---|
| Model | DS-161 |
| Throughput | 200 tests/hour |
| Assay Type | Random access with direct reading system |
| Methodology | End point, Kinetic, Fixed time, Immunoturbidimetry, Single wavelength, Bichromatic, Multi-standard, Reagent blank, Serum blank etc. |
| Sample Volume | 1-100ul, 0.1ul/step |
| Reagent Volume | 1-500ul, 0.1ul/step |
| Sample Position | 40 positions. Including standard, control and STAT, either serum cup or primary tube available |
| Reagent Position | 42 positions: R1:24, R2:18 |
| Refrigeration System | Keep reagent chamber at 2-8℃ |
| Mixing System | One independent mixer |
| Reagent/Sample Probe | With liquid level sensor, collision sensor, automatically detect reagent residue in reagent bottle |
| Reaction Time | 10 minutes |
| Cuvette | 60 high quality UV-transmitted plastic cuvettes, and quartz glass cuvettes available on request |
| Washing | 5 channels automatic washing cuvette system; waste liquid collected by high concentration and low concentration respectively; automatically dry cuvette |
| Alarm | Automatic alarm for malfunction(collision, reagent/sample insufficiency, full waste container etc.) |
| Standard | Linear, Non-linear, Multi standard, K factor |
| Sample Dilution/Retest | Sample can be diluted and retested if result is out of range or sample is insufficient |
| Light Source | Halogen lamp |
| Wavelength | 330nm-850nm; standard equipped with 8 filters, 2 more filters available on request |
| Absorbance Range | 0~3.5Abs |
| Resolution | 0.0001Abs |
| Absorbance Drift | ≤0.002 |
| Temperature Drift | ±0.1℃ |
| Carry Over | ≤0.01% |
| Precision | 0.0001O.D. |
| Auto Blank Function | Auto zero setting and deduct/record blank value for each cuvette to ensure accurate result |
| Emergency Test | Emergency sample test prior |
| Reliable Accuracy | Equipped with automatically test error diagnosis and correcting, automatic maintenance etc. |
| Control/Calibration | With patented auto control and calibration management function |
| Data Storage | More than 5000 items and 2 million results can be edited and stored |
| Software | Stronger and more convenient software with multiple languages |
| Printer | Multiple report formats available |
| Power Supply | 220V or 110V, 50Hz or 60Hz |
| Dimension | 65×53×52(cm) |