Technical Parameter:
| Item | HB7021 |
|---|---|
| Model | HB7021 |
| Automatic | Fully automatic |
| Throughput | 60 tests / hr |
| Assay Items | 22 parameters, 3 parts differential of WBC |
| Sample Volume | 13ul (whole blood), 20ul (prediluted blood) |
| Measuring Principle | Electro-impedance, photometric analysis |
| Parameter | WBC, LY, LY%, MO, MO%, GR, GR%RBC, HGB, HCT, MCV, RDW-SD, RDW-CV, MCH, MCHCPLT, MPV, PCT, PDWWBC histograms, RBC histograms, PLT histograms |
| Linearity Range | WBC: 0.0-99.9 (×10⁹/L)RBC: 0.0-9.99 (×10¹²/L)HGB: 0.0-300 (g/L)PLT:0-999 (×10⁹/L) |
| Accuracy (CV %) | WBC≤2.0% RBC≤1.5% PCT≤3.0%PLT ≤5.0% HGB ≤1.5% MPV ≤3.0%HCT ≤ 2.0% MCV≤1.0% PDW≤ 3.0%MCH ≤ 2.0%, MCHC≤2.0% |
| Display | LCD |
| Storage | 200,000 results including histograms |
| Output | One RS232, one PS/2 and two USB. Available for LIS |
| Printer | Built-in thermal printer, USB to external printer(optional) |
| Data Input | Keypad(mouse and keyboard available) |
| QC | Automatically calculates X̄,SD,CV% and saves data |
| Language | English and Spanish.Other language on request |
| Power Supply | AC110V 60HZ or 220V 50HZ |
| Dimension | 43cm32cm50.5cm |
| Weight | 20.00 kg |