Technical Specification
Anaesthesia main unit
High luminance digital LED screen
Close,semi-close,semi-open system
Application: Adult,Pediatric
Low gas alarming: Low O2 gas pressure,it will gives alarming more than 7 seconds
High pressure limitation: Whenever airway pressure exceed the pre-set high pressure,
ventilator offers protection automatically,to assure airway pressure can not increase
anymore which supplies safety to patient
Power failure alarming,power extendable function
O2 flush:25~75L/min
Gas Pressure: O2: 0.28~0.6MPa
Power:AC 220V±10%,50Hz±1%
Flow Meter
O2/N2O-2 tubes(0.05~1L/min,1.1~10L/min)
Anaesthesia Ventilator
Adult bellow 1set,Optional 1 set Pediatric bellow
Pneumatic driven electric control
Monitor parameters: Tidal volume,Respiratory Frequency,I:E rate,Peak airway pressure
Working mode: Manual,Auto
Tidal volume: 0,50~1500ml
Respiratory frequency:4~60bpm
I/E Ratio: 1:1~1:4 adjustable
Synchronizer trigger pressure:-3~0KPa
High pressure limitation:2~6KPa
Low pressure limitation:0.6~2KPa
Respiratory Circuit
Exterior CO2 absorber,simply disassembly,convenient to replacing “soda lime”.
Vaporizer
One precise vaporizer(Enflurane、Halothane、Isoflurane、Sevoflurane)
Adjustable range: 0.5~5%Vol
Automatic compensation function –flow,pressure,temperature
Self-lock function
Volume:180ml(Optional:250ml)