Adopting a dual circuit design, when one circuit fails, it can be switched to the other circuit; It can achieve non-stop gas maintenance without affecting pressure, meeting the requirement of "continuous gas supply under a single fault" stipulated in GB 50751-2013 "Technical Specification for Medical Gas Engineering". And provide the testing report of GB 50751-2012 "Technical Specification for Medical Gas Engineering"
The pipeline adopts a design of one backup and one use, which is convenient for maintenance; | |
Box size | 420mm * 450mm * 170mm |
Import pressure P1 | 0.6~0.8MPa |
Export pressure P2 | 0.35~0.6MPa (adjustable) |
Precision pressure gauge with safety valve | |
Medium | Air; Flow rate ≥ 300L/min |
Static characteristics | P2 pressure increased by 0.06MPa |
Brass pressure regulator, brass maintenance valve | |
Degreasing seamless stainless steel 304 pipeline | |